Paul Little aCommunity
Clinical Sciences Division (Primary Medical Care Group), Faculty of
Medicine, Health and Biological Sciences, Southampton University,
Aldermoor Health Centre, Southampton SO16 5ST, b Nightingale
Surgery, Greatwell Drive, Romsey SO51 7QN, c St Clements Surgery,
Winchester SO23 8AD Correspondence
to: P Little psl3{at}soton.ac.uk
Primary care p 258
| |
Abstract |
|---|
|
|
|---|
Objective:
To assess alternatives to measuring
ambulatory pressure, which best predicts response to treatment and
adverse outcome.
Setting:
Three general practices in England.
Design:
Validation study.
Participants:
Patients with newly diagnosed high or
borderline high blood pressure; patients receiving treatment for
hypertension but with poor control.
Main outcome measures:
Overall agreement with
ambulatory pressure; prediction of high ambulatory pressure (>135/85
mm Hg) and treatment thresholds.
Results:
Readings made by doctors were much higher than ambulatory systolic pressure (difference 18.9 mm Hg, 95% confidence interval 16.1 to 21.7), as were recent readings made in the
clinic outside research settings (19.9 mm Hg,17.6 to 22.1). This
applied equally to treated patients with poor control (doctor v ambulatory 21.4 mm Hg, 17.3 to 25.4). Doctors' and recent
clinic readings ranked systolic pressure poorly compared with
ambulatory pressure and other measurements (doctor r=0.46;
clinic 0.47; repeated readings by nurse 0.60; repeated self measurement
0.73; home readings 0.75) and were not specific at predicting high
blood pressure (doctor 26%; recent clinic 15%; nurse 72%; patient in
surgery 81%; home 60%), with poor likelihood ratios for a positive
test (doctor 1.2; clinic 1.1; nurse 2.1, patient in surgery 4.7; home 2.2). Nor were doctor or recent clinic measures specific in predicting treatment thresholds.
Conclusion:
The "white coat" effect is important
in diagnosing and assessing control of hypertension in primary care and
is not a research artefact. If ambulatory or home measurements are not available, repeated measurements by the nurse or patient should result
in considerably less unnecessary monitoring, initiation, or changing of
treatment. It is time to stop using high blood pressure readings
documented by general practitioners to make treatment decisions.
|
What is already known on this topic
Preliminary evidence suggests that measurements by doctors are likely to be higher than those made by nurses, technicians, or patients No study has compared all the available measures in a typical primary care setting with ambulatory blood pressure in patients with newly diagnosed and established hypertension What this study adds
If ambulatory measurement is not possible, repeated measurement by a nurse or by the patient will result in much less unnecessary treatment or change in treatment for high blood pressure |
| |
Introduction |
|---|
|
|
|---|
Hypertension is perhaps the most common reason for initiation of lifelong drug treatment and ongoing management by doctors. Six prospective studies have shown that ambulatory blood pressures may be a much better predictor of target organ damage and subsequent adverse events than measurements made in a clinic.1 As these results were found in research studies and mostly not in typical primary care settings, however, patients may have had a higher "alerting response" than in everyday settings with their family doctor or nurse. It is thus important to clarify whether the white coat effect applies equally outside a research study and in typical family practice settings.
Why is ambulatory monitoring not commonly used to make management decisions? The problem is not just extrapolating results from research or secondary care to routine settings but that clinic derived thresholds have been used in previous research to make treatment decisions. However, several lines of evidence show that patients with daytime ambulatory pressure lower than 135/85 mm Hg have a low risk of subsequent events.2 An ambulatory pressure of 135/85 mm Hg thus represents good control and approximately corresponds to a clinic pressure of 140/90 mm Hg,2 a generally accepted marker for control.3 The threshold for diagnosis in the clinic is usually higher (>160/100 mm Hg for most patients),4 so a higher ambulatory threshold of 145/95 mm Hg has been proposed.5 Recent guidelines recommended ambulatory monitoring for both initial diagnosis and assessing control,6 although few studies have looked at the assessment role in primary care. One trial in a mixed setting showed that management according to ambulatory pressure resulted in fewer visits, less use of drugs, and similar final blood pressures.7 Further evidence is needed from typical primary care settings to explore the implications of using ambulatory pressures and other alternatives, both in the initiation of treatment and in monitoring control.
What about other alternatives? Preliminary evidence, mostly from other settings, indicates that measurements by a nurse or technician, repeated measurements, or home measurements may be closer to ambulatory pressure.7-14 To our knowledge, no study in a typical primary care setting has compared these methods with ambulatory monitoring. Another alternative is self measurement by patients with equipment in the clinic, which to our knowledge has never been assessed.
We set out to assess the following in typical primary care settings.
(1) The white coat effect in the broad group in which decisions are
usually made on clinic readings (diagnosis and monitoring control). (2)
The extent of the white coat effect as a research artefact. (3) The
agreement
ranking, mean difference, and detection of high
pressures
between ambulatory monitoring of blood pressure and the
realistic alternatives (measurement by doctor or nurse, self
measurement in surgery, home). (4) The potential implications of using
alternative methods of blood pressure measurement in predicting
treatment thresholds.
| |
Methods |
|---|
|
|
|---|
Setting
Eight doctors and three practice nurses from three varied
practices (deprived urban, cathedral city, market town), each serving
8000 patients, agreed to participate. The doctors and nurses had not
previously been involved in research on hypertension. Practice nurses
were trained to use the 24 hour monitor and to follow the protocol of measurements.
Participants
Two hundred patients in whom management changes were being
considered on the basis of clinic readings participated in the study.
Participants were in two categories: newly diagnosed or borderline
hypertension (three clinic readings of systolic blood pressure >140 mm
Hg or diastolic pressure >90 mm Hg),4 or established
hypertension (three clinic readings >160/100 mm Hg) being treated but
with poor control (>140/90 mm Hg).3
Most participants were referred opportunistically by doctors; nine eligible patients declined to participate. In one practice some patients were also invited on the basis of high readings documented in the notes.
Equipment
We used the UA-751421 semiautomated device and the OMRON HEM 705CP
for measurements by a nurse and home readings by
patients.
15 16
For 24 hour readings we used the
TM-2421,17 which is a later, better validated version of
the extensively used and validated TM-2420. We also assessed the
calibrated mercury sphygmomanometers in current use by the practices.
Large cuffs were used where appropriate.
Sample size
Assuming that 50% of patients have high systolic ambulatory
pressure and that other methods can detect this with a sensitivity and
specificity of 70%, with a 95% confidence interval around this
estimate of 20% (plus or minus 10%), then we needed 180 participants.
Blood pressure measurements
The order of measurement reflected the way measurements might be
used in practice. Firstly, two measurements were made by the nurse (two
visits). The nurse took readings with two types of device three times
each, sequentially at every visit, after the patient had been seated
for five minutes: a calibrated mercury sphygmomanometer and a
validated, calibrated semiautomated cuff oscillometric device. The
nurse alternated the two devices and the arm used; which device was
used first was randomised. Secondly, the patient carried out self
measurement in the surgery (the last 70 patients were invited; 59 came
once and 52 twice). A room was provided for the patients to come and
take three measurements with the semiautomated device, having
previously been instructed by the nurse. Thirdly, 24 hour ambulatory
measurement readings were taken at half hourly intervals during the day
(0700-2300) and hourly at night (2300-0700). Fourthly, patients took
readings at home using semiautomated machines. Patients took four
readings a day (two in the morning and two in the evening) on
successive days until 14 readings had been taken (which enabled cross
validation with the machines' 14 reading memory). Fifthly, after all
the other measurements had been carried out, the doctor measured blood pressure with a calibrated mercury sphygmomanometer three times sequentially after the patient had been seated for five minutes.
We randomised the order of home and ambulatory measurements. For the first 130 patients ambulatory or home monitoring took place between the first and second visit to the nurse, with the other measurement (whichever of home and ambulatory had not been done) after the second visit. To estimate whether the fall in blood pressure between the two nurse visits was due to habituation to the intervening ambulatory or home measurements, for the last 70 patients both the home and ambulatory measurements took place after both nurse visits and self measurement in the surgery.
We also recorded up to three recent measurements made in the clinic and recorded in the notes before participation in the study. By their nature, these readings were not controlled in terms of a study protocol, and most readings were made by doctors in the weeks immediately before the study.
Data entry and analysis
We used SPSS and Stata for Windows to set up and analyse the
database. We assessed whether the readings were significantly different
by using repeated measures analysis of variance and the Bonferroni
correction for retrospective comparisons. We made several comparisons
of each "test" method and ambulatory monitoring: mean difference
(standard deviation; 95% confidence intervals); rank correlation
from the Bland Altman plot (plot of the difference between measurements
against the mean); rank correlation from a simple scatter plot to
document how each test method ranked the blood pressures and thus the
risk due to blood pressure; the sensitivity, specificity, and
likelihood ratios for a positive and negative test, for predicting high
blood pressure and treatment thresholds.
| |
Results |
|---|
|
|
|---|
Sample
Of the 200 participants, 107 (54%) were women, 63/194 (33%) were
over the age of 65, 96 had newly diagnosed high or borderline high
blood pressure (systolic pressure 161 (SD 16) mm Hg; diastolic pressure
95 (9) mm Hg), and 104 were treated but had poor control (systolic
pressure 163 (16) mm Hg; diastolic pressure 94 (9) mm Hg). Six patients
did not have ambulatory readings (withdrawn or cuff uncomfortable); 173 patients had a complete set of all
readings.
|
|
Preliminary analysis: are different readings significantly
different?
Repeated measures analysis of variance (excluding self measurement
by patients in the surgery, owing to the small numbers) showed that
measurements of systolic pressure were not equivalent
(F=63.1, P<0.001). Post hoc tests showed that all the means differed significantly from ambulatory pressure and from each
other, except readings by the doctor and recent clinic readings (which
did not differ significantly from each other) and home readings and the
second measurement by the nurse (which did not differ significantly
from each other). Significant differences also existed between measures
of diastolic blood pressure (F=74.4, P<0.001). In post
hoc analysis diastolic ambulatory pressure differed significantly from
all other measures, and readings by the doctors and clinic readings
differed significantly from all other measures but not from each other.
Estimating agreement: main results
Figures 1 and 2 illustrate the results in tables 1, 2 and 3
graphically by taking data from the first line of tables 1 and 2.
Overall agreement for systolic pressure is illustrated by the scatter
plot (fig 1), with a moderate rank correlation (r=0.46);
the cut-off points show that readings by the doctor are reasonably
sensitive in detecting high ambulatory pressure, but poorly specific.
The Bland Altman plot (fig 2) is summarised by the mean difference (the
readings by the doctor on average exceed ambulatory pressure by 18.9 mm
Hg) and the positive rank correlation (Spearman's
r=0.32). Thus the difference between readings by the
doctor and ambulatory pressure increases as the blood pressure
increases.
|
|
|
Readings made by the doctor were much higher than systolic ambulatory pressure (difference=18.9 mm Hg, 95% confidence interval 16.1 to 21.7), as were recent clinic readings not made in a research study (19.9 mm Hg, 17.6 to 22.1). The white coat effect applied equally for patients on established treatment with poor control (readings by doctor v ambulatory pressure, difference=21.4 mm Hg, 17.3 to 25.4). For most methods the difference from ambulatory monitoring increased with blood pressure (a positive correlation on the Bland Altman plots, see table 1). Readings by the doctor and in the clinic also ranked systolic ambulatory pressure poorly compared with other methods (table 1). Most methods were sensitive in predicting high systolic ambulatory pressure (all sensitivity >75%), but readings by the doctor and recent clinic readings were not specific, with poor likelihood ratios (table 2). Nor were readings by the doctor or recent clinic readings specific in predicting ambulatory systolic treatment thresholds, having poor likelihood ratios (table 3). Measurement by the doctor and recent clinic readings performed slightly better for diastolic pressures than for systolic pressures, although other methods still performed better, with higher likelihood ratios for a positive test.
Reliability of ambulatory pressure
Five consecutive patients had ambulatory monitoring repeated after
two weeks, which showed good agreement (r=0.86 for daytime
systolic pressure, r=0.84 for 24 hour systolic pressure; mean differences
2.2 (SD 5.9) mm Hg and
3 (4.8) mm Hg). This confirms previous evidence about the reliability of ambulatory pressure.18
Attenuation of white coat effect with visits to nurse
The fall in blood pressure between the two visits to the nurse was
even greater for the last 70 patients, in whom no intervening
measurements occurred. For these 70 patients the difference between
ambulatory systolic pressure and measurements taken by the nurse was 11 mm Hg for the first visit and 2.8 mm Hg for the second visit.
Patient documentation at home compared with semiautomated
sphygmomanometer memory
Agreement existed for consecutive patients between the 14 readings
documented by patients and the machine memory for both systolic
pressure (n=21; r=0.97; mean difference 3.5 (SD 9.3) mm
Hg) and diastolic pressure (r=0.85; mean difference 1.4 (6.0) mm Hg). The differences between readings recorded by the patient
and the machine reflect the fact that "practice" readings or
readings taken at other times of day (for example, during work) were
also recorded by the machine.
| |
Discussion |
|---|
|
|
|---|
Limitations of the study
Routine equipment
The agreement with ambulatory
pressure was similar when routine equipment was used and when well
validated equipment was used. Thus the main limitation of measurement
in routine clinic settings is not the type of sphygmomanometer used.
We used patients with newly diagnosed
hypertension and patients with poorly controlled hypertension, but this
reflected recent guidance.6 Furthermore, the subgroups
(newly diagnosed and established hypertension) had very similar blood pressures.
Arbitrary cut-off points to define potential management
decisions
Exact management thresholds will always be debated.
Nevertheless, agreement exists that poor control in clinic readings for
most patients is >140/90 mm Hg,3 and that for ambulatory
readings
135/85 mm Hg represents good control (although patients
with additional risk factors may need tighter
control).
2 5 6
There is also agreement that patients
with clinic pressures of >160/100 mm Hg need drug treatment to be
started,4 but debate continues about the threshold for
ambulatory pressure. We used a level defined by previous
recommendations (>145/95 mm Hg, which corresponds approximately to the
160/100 mm Hg for clinic pressures).5 Other authors have
suggested using a similar criterion by adding 10 mm Hg systolic and 5 mm Hg diastolic to home or ambulatory pressures to provide
"equivalent" clinic readings to inform decision making.14 The poor specificity and likelihood ratios of
readings by the doctor and in the clinic with either a single cut-off
point for high pressure or different cut-off points for treatment
suggests that the particular thresholds chosen do not alter the
inferences from the study.
Order of measurements
Recent measurements in the
clinic (that is, historic measurements) were very similar to
measurements by the doctor (the last recorded in the study).
Furthermore, the drop between the two measurements by the nurse for the
first 130 patients could not be explained by the intervening home or 24 hour measurements, as a similar or greater effect was seen in the last
70 patients, for whom there were no intervening measurements. Thus the
differences between measurements are likely to be due not to effects of
order or time but to the different alerting responses for different measurements.
The white coat effect
artefact of setting, sample, or research
studies?
The estimates of the white coat effect in this study are similar
to those in previous work, mostly not from typical primary care
settings.8-14 This was true for both patients with newly
diagnosed hypertension and those with established hypertension, and for
both measurements taken as part of the study and recent measurements
documented in the clinical notes. Thus the white coat effect observed
in the major prognostic studies to date is not likely to be an artefact
of other settings or of research studies and applies equally to new
diagnosis and assessment of control.
Although debate continues about just how benign white coat hypertension
is,19 such studies have been criticised as some have not
measured the white coat effect in a standard way.20 Prospective studies indicate that patients with white coat hypertension are at considerably reduced risk compared with those with higher ambulatory pressures and that treatment modifies blood pressure and
outcomes only in patients with high ambulatory blood
pressure.
2 7 19
The overzealous initiation and
maintenance of treatment for white coat hypertension represents an
enormous opportunity cost for health professionals and for patients, in
addition to the associated iatrogenesis
particularly unnecessary
anxiety21 and side effects.
22 23
Are the alternatives to measurements by a doctor better?
Evidence exists from prospective studies about the relation
between ambulatory blood pressure and outcome.1 Nevertheless, ambulatory monitoring equipment is expensive (the machine
used in this study cost £2000), and as approximately half of all
patients with hypertension are poorly controlled each practice would
need several machines to cope with the workload. Although more
automated downloading of results should potentially reduce costs, staff
costs (setting up, reviewing, downloading, printing) are also currently
high, meaning that costs are not reduced
overall.7
This study shows that
repeated readings by a nurse in primary care provide a better
assessment than readings by doctors, supporting research from other
settings.10 However, staff costs are similar to those for
ambulatory monitoring.7
Home measurement
This study confirms that
patients can accurately measure and record their blood pressure
themselves at home,
10 14 24
with great potential
advantages of lower equipment and staff costs compared with ambulatory
pressure. Although ambulatory monitoring can assess night time dipping
and hence help with diagnosis of secondary hypertension, this can only
be securely diagnosed by investigation (renal function tests,
ultrasonography, electrolytes, vanillylmandelic acid).
Self measurement by patients in the clinic
Self
measurement by patients in the clinic may provide similar levels of
overall agreement with ambulatory pressure to home measurement. The
advantages are that practice staff are available for questions and that
equipment costs are minimal (one machine can be kept in the practice,
rather than several machines being lent out). The disadvantages are
that more visits to the clinic are needed than for home readings and
that suitable rooms have to be made available. The estimates from this study are also less precise and need confirmation (only 52 of the 70 patients invited completed both measurements).
Conclusion
The "white coat" effect is important in diagnosing and
assessing control of hypertension in primary care and is not a research
artefact. If ambulatory or home measurements are not available,
repeated measurements by a nurse or the patient should result in
considerably less unnecessary monitoring, initiation, and changing of
treatment. It is time to stop using high blood pressure readings
documented by general practitioners to make decisions about treatment.
| |
Acknowledgments |
|---|
We thank the doctors, staff, and patients at Aldermoor Health Centre, Nightingale Surgery, and St Clements Surgery. We also thank Eoin O'Brien for advice and helpful comments. DM is now employed at the Division of Public Health and Primary Care, Institute of Health Sciences, Oxford OX3 7LF.
Contributors: PL and AFB had the original ideas for the study. All authors contributed to the development of the protocol. JB, JM, and LB ran the study on a day to day basis. PL and LB performed the analysis. All authors contributed to writing the paper. PL is the guarantor of the study.
| |
Footnotes |
|---|
Funding: NHS Regional Research and Development Grant and the HOPE charity. PL is funded by the Medical Research Council.
Competing interests: None declared.
| |
References |
|---|
|
|
|---|
| 1. | Pickering TG. Ambulatory blood pressure monitoring. Curr Hypertens Rep 2000; 2: 558-564[Medline]. |
| 2. | Pickering TG. What is the "normal" 24 h, awake, and asleep blood pressure? Blood Press Monit 1999; 4(suppl 2): S3-S7. |
| 3. |
O'Rorke J, Scott Richardson W.
Evidence based management of hypertension: what to do when blood pressure is difficult to control.
BMJ
2001;
322:
1229-1232 |
| 4. | Ramsay L, Williams B, Johnson G, MacGregor GA, Poston L, Potter J, et al. Guidelines for management of hypertension: report of the third working party of the British Hypertension Society. J Hum Hypertens 1999; 13: 569-592[CrossRef][Web of Science][Medline]. |
| 5. |
Prasad N, Isles C.
Ambulatory blood pressure monitoring: a guide for general practitioners.
BMJ
1996;
313:
1535-1541 |
| 6. |
O'Brien E, Coats A, Owens P, Petrie J, Padfield P, Littler W, et al.
Use and interpretation of ambulatory blood pressure monitoring: recommendations of the British Hypertension Society.
BMJ
2000;
320:
1128-1134 |
| 7. |
Staessen JA, Byttebier G, Buntinx F, Celis H, O'Brien ET, Fagard R.
Antihypertensive treatment based on conventional or ambulatory blood pressure measurement: a randomized controlled trial.
JAMA
1997;
278:
1065-1072 |
| 8. | Pickering TG. Blood pressure measurement and detection of hypertension. Lancet 1994; 344: 31-35[CrossRef][Web of Science][Medline]. |
| 9. | Cox JP, O'Malley K, O'Brien E. Ambulatory blood pressure measurement in general practice. Br J Gen Pract 1992; 42: 402-403[Web of Science][Medline]. |
| 10. | Brueren M, Shouten H, Leeuw P, Montfrans G, Van Ree J. A series of self measurements by the patient is a reliable alternative to ambulatory blood pressure measurement. Br J Gen Pract 1998; 48: 1585-1589[Web of Science][Medline]. |
| 11. | Veerman DP, Van Montfrans GA. Nurse-measured or ambulatory blood pressure in routine hypertension care. J Hypertens 1993; 11: 287-292[CrossRef][Web of Science][Medline]. |
| 12. | Veerman DP, de Blok K, Delemarre B, Van Montfrans GA. Office, nurse, basal and ambulatory blood pressure as predictors of hypertensive target organ damage in male and female patients. J Hum Hypertens 1996; 10: 9-15[Web of Science][Medline]. |
| 13. | Fagard R, Staessen J, Thijs L, Amery A. Multiple standardized clinic blood pressures may predict left ventricular mass as well as ambulatory monitoring: a metaanalysis of comparative studies. Am J Hypertens 1995; 8: 533-540[CrossRef][Web of Science][Medline]. |
| 14. |
Aylett M.
Ambulatory or self blood pressure measurement? Improving the diagnosis of hypertension.
Fam Pract
1994;
11:
197-200 |
| 15. | Jamieson MJ, Webster J, Witte K, Huggins M, MacDonald TM, de Beaux A, et al. An evaluation of the A&D UA-751 semi-automated cuff-oscillometric sphygmomanometer. J Hypertens 1990; 8: 377-381[CrossRef][Web of Science][Medline]. |
| 16. | O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure according to the revised BHS protocol: the OMRON HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Press Monit 1996; 1: 55-61[Medline]. |
| 17. | Imai Y, Sasaki S, Minami N, Munakata M. The accuracy and performance of the A&D TM 2421, a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method. Am J Hypertens 1992; 5: 719-726[Web of Science][Medline]. |
| 18. |
James GD, Pickering TG, Yee LS, Harshfield GA.
The reproducibility of average ambulatory home and clinic pressures.
Hypertension
1988;
11:
545-549 |
| 19. | McGrath B. Is white coat hypertension innocent? Lancet 1996; 348: 630[CrossRef][Web of Science][Medline]. |
| 20. |
Muscholl M, Hense H-W, Brockel U, Doring A, Riegger G.
Changes in left ventricular structure and function in patients with white coat hypertension: cross sectional survey.
BMJ
1998;
317:
565-570 |
| 21. | Marteau T. Reducing the psychological costs. BMJ 1991; 301: 26-28[Web of Science]. |
| 22. | Medical Research Council Working Party on Mild Hypertension. Adverse reaction to propranolol and bendrofluazide for the treatment of mild hypertension. Lancet 1981; 2: 539-543[Medline]. |
| 23. |
Finnerty FA.
Step down therapy for hypertension: importance in long-term management.
JAMA
1981;
246:
2593-2596 |
| 24. | Aylett M, Marples G, Jones K. Home blood pressure monitoring: its effect on the management of hypertension in general practice. Br J Gen Pract 1999; 49: 725-728[Web of Science][Medline]. |
(Accepted 21 March 2002)
Read all Rapid Responses